Medtronic | Company Profile

← Back to Biosensors & POCT guides

COMPANY PROFILE / BIOSENSORS

MiniMed (Medtronic)
Medtronic's former diabetes business, listed in March 2026, which sells insulin pumps together with continuous glucose monitoring (CGM) and uses both its own sensors and sensors made by Abbott

MiniMed comes up in any discussion of biosensors because it sells automated insulin delivery systems that take a CGM sensor's readings as input and adjust insulin dosing automatically. It used to be the Diabetes business of Medtronic plc, but as MiniMed Group, Inc. it listed on Nasdaq on 9 March 2026 (ticker MMED). Medtronic holds about 90%, and on 14 September 2026 it launched an exchange offer that lets its own shareholders swap Medtronic shares for MiniMed shares. For the fiscal year ended 24 April 2026 (FY2026) MiniMed reported sales of US$3,102 million, R&D spending of US$448 million and about 8,000 employees. This profile sets out, from primary sources only, what the company makes, what it invests in, who it works with and what challenges it faces. It is not a recommendation to use any product.

Sources: MiniMed Group, Inc.'s Form 10-K (FY2026), Form 10-Q (May to July 2026), results release and IPO 8-K; Medtronic plc's Form 10-K and its exchange-offer announcement and 8-K; FDA PMA and 510(k) documents and the recall database; Abbott's announcement; and ClinicalTrials.gov, used as primary material. Last updated October 2026.

What this article covers
  1. MiniMed in 30 seconds, and where it stands in biosensors
  2. The logo and how the name is written
  3. Where the company sits in the field
  4. Scale
  5. Resources committed to biosensors
  6. What it has done so far
  7. What comes next
  8. What the company is aiming for in biosensors
  9. Map of partnerships in biosensors
  10. Technologies, products and services
  11. The risks this company carries
  12. Glossary, references and claim-to-source audit

1. MiniMed in 30 seconds, and where it stands in biosensors

LEGAL NAMEMiniMed Group, Inc.Formerly Medtronic's Diabetes business. A Delaware corporation; one operating entity is Medtronic MiniMed, Inc.
HEADQUARTERSNorthridge, California, U.S.Two main manufacturing sites: California and Puerto Rico
LISTING AND PARENTNasdaq (MMED)
IPO on 9 March 2026
Medtronic plc (Irish company, NYSE: MDT) holds about 90%
CATEGORY IN THIS FIELDDiabetes & continuous monitoring (primary)A company that sells pump and CGM as a system, not a sensor specialist
SALES (FY2026)US$3,102 millionYear to 24 April 2026. US$2,715 million the year before
R&D EXPENSE (FY2026)US$448 millionUS$436 million the year before. 10-K figures
EMPLOYEESAbout 8,000As of April 2026; over 2,800 of them commercial staff (10-K)
CGM SENSORSSimplera Sync, Guardian 4 (own)
Instinct (made by Abbott)
The 10-K says it uses its own and Abbott-made sensors
The one thing to grasp first

MiniMed is best understood not as "a sensor company" but as "a company that sells automated insulin delivery systems in which the sensor is one component". About 82% of FY2026 sales came from recurring revenue such as CGM, consumables, software and services, according to the 10-K Sourced. Its sensors are both its own (Simplera Sync, Guardian 4) and Instinct, made by Abbott under a supply and integration agreement. Also, Medtronic still holds about 90% after the IPO, and what happens next depends on the outcome of the exchange offer, which is scheduled to expire on 9 October 2026. As of 4 October 2026 no result had been announced Not yet confirmed.

2. The logo and how the name is written

A note on names: the company that listed in March 2026 is MiniMed Group, Inc. Before the separation it was the Diabetes business of Medtronic plc, and the earlier product name was Medtronic MiniMed (the applicant name in FDA documents is still Medtronic MiniMed, Inc.). The title follows the series convention of naming the operating business with its well-known parent in parentheses, "MiniMed (Medtronic)". The folder name stays Medtronic. MiniMed 780G, MiniMed Flex, Simplera Sync, Guardian 4 and Instinct are product names, not companies. The logo was taken from the company's official website (minimed.com).

Official site: https://www.minimed.com/

3. Where the company sits in the field

Four types of company involved in biosensors Even among "sensor companies", where they measure, who buys and how heavily they are regulated differ widely Diabetes & continuous monitoring In vitro diagnostics & POCT Wearables & consumer devices Materials & components Continuous glucose sensing Analyzers and reagents Daily sensing in devices Enzymes and electrodes Dexcom / Abbott Medtronic / Senseonics Roche / bioMérieux Siemens Healthineers Apple / Samsung Oura and others Toyobo / Paragraf and others Sensors and clinical data Tied to rules and payment Large installed base Tied to test demand Many users, wide reach Limited by medical claims Quality and consistency Tied to device demand MiniMed is mainly diabetes and continuous monitoring: insulin pumps and CGM sold as one system Fiscal 2026 sales were US$3,102 million; its own Simplera sensors are used alongside Abbott-made ones
Fig. 1 The four types of company involved in biosensors, and where MiniMed sits. The classification is ours, and the company names are examples drawn from the companies covered in this series Our calculation. Simplera and Instinct are product names, not companies.

4. Scale

MiniMed is a diabetes-only company, so company-wide figures are the diabetes figures. Its parent Medtronic is large, so group-wide figures are kept brief. Each table comes from the company's own 10-K Sourced.

MiniMed Group (US$ million)FY2025FY2026Change
Sales2,7153,102+14% (10-K)
Gross profit1,5281,680+10%
R&D expense436448+3%
Operating income (loss)-146-190Loss widened
Net income (loss)-198-317Loss widened
Purchases of property and equipment193223FY2024: 148
Medtronic plc (whole group)FY2025FY2026Change
Group sales (US$ million)33,53736,364+8%
Of which the Diabetes segment2,7553,112+13%

Medtronic's Diabetes segment is about 8.6% of group sales Our calculation (3,112 / 36,364). MiniMed's own 10-K sales (3,102) and Medtronic's segment sales (3,112) differ by US$10 million; the reason was not confirmed here. About 70% of MiniMed's sales are outside the U.S. (10-K, FY2026). For May to July 2026 (the first quarter of FY2027), sales were US$843 million (up 16.6% on the year), with US$240 million in the U.S. and US$603 million elsewhere. Operating income was US$5 million (against a US$13 million loss a year earlier) (10-Q and results release) Sourced.

A caution

The first quarter of FY2027 includes roughly one extra week of sales because of the 52/53-week fiscal calendar (the company says about 4 to 6 points of growth). Comparing it directly with the prior-year quarter makes growth look larger. MiniMed's financials also cover the period when it was a division of Medtronic (transfers with the parent are recorded), so cost structures before and after listing may differ and year-to-year comparisons need care Not yet confirmed.

5. Resources committed to biosensors

ResourceWhat primary sources confirmStatus
PeopleAbout 8,000 employees (April 2026), over 2,800 of them commercial staff. Two manufacturing sites (California and Puerto Rico). More than 2,000 patents and patent applications as of May 2026 (10-K). Headcount for sensor R&D alone is not disclosedSourced / partly undisclosed
Money (R&D)US$436 million in FY2025 and US$448 million in FY2026. For May to July 2026: US$115 million (US$125 million a year earlier). R&D for sensors alone is not disclosedSourced / partly undisclosed
Money (capital expenditure)Purchases of property and equipment: US$148 million in FY2024, US$193 million in FY2025 and US$223 million in FY2026. The results release names Simplera (the company's own sensor) capacity expansion as a driver of growth outside the U.S.; the breakdown of the investment was not confirmedSourced / breakdown not confirmed
External funding and investment(1) IPO: 28,000,000 shares sold at US$20.00 on 9 March 2026 (8-K); cash from the share issuance was US$538 million (10-K cash flow statement). (2) Blackstone: under co-development agreements signed by Medtronic, the company received funding for specified diabetes products (next-generation MiniMed Flex pump, MiniMed Fit patch pump and others) and owes royalties. On FDA clearance of MiniMed Flex it recorded a one-time charge of US$157 million in the fourth quarter of FY2026 for minimum royalty obligations; a termination payment can be up to US$216 million per agreement. These agreements were assigned to MiniMed in the separationSourced
Acquisitions and joint venturesNo acquisition or joint venture in sensors was found in the sources checked. There is, however, the supply and integration agreement with Abbott (below)Not confirmed
Universities, research institutes, public fundingClinicalTrials.gov lists 79 studies with "Medtronic MiniMed" as sponsor or collaborator (searched 4 October 2026). NIH or other public grant funding is not confirmed in the 10-KSourced / public funding not confirmed

Regulatory standing (confirmed in FDA's official documents)

ProductUse (as written in FDA documents)Category and numberDate
MiniMed 780G system (including the Guardian 4 sensor)A system combining insulin delivery and CGM. Changes to the SmartGuard technology and an expansion of the indications to include the Guardian 4 sensorPMA supplement approval P160017/S091 (original approval: MiniMed 670G, P160017, 28 September 2016)21 April 2023
Simplera system (CGM)Real-time CGM; at approval, for ages 18 and older, intended to replace fingerstick testingPMA supplement approval P160007/S047 (original approval: Guardian Connect, 8 March 2018)24 July 2024
Simplera Sync sensor and Guardian 4 sensorIntended to function as components of a Medtronic automated insulin dosing systemPMA approval P250012 (product code SFI)6 August 2025
SmartGuard technology (automated dosing software)For use with compatible iCGMs, compatible Medtronic CGMs and alternate controller enabled (ACE) pumps. Type 1 diabetes aged 7 and older; type 2 diabetes aged 18 and older requiring insulin. Prescription required. The summary records integration testing with an iCGM, "Instinct, made by Abbott"510(k) clearance K253585 (product code QJI)14 January 2026
MiniMed 780G insulin pumpSubcutaneous insulin delivery, aged 7 and older, prescription required; communicates securely with external automated insulin dosing software510(k) clearance K253470 (product code QFG)15 January 2026
MiniMed Flex pumpAn alternate controller enabled (ACE) insulin pump that communicates with compatible iCGMs, Medtronic CGMs and automated dosing software. Launched in the U.S. in June 2026 (10-K)510(k) clearance K253743 (product code QFG)12 March 2026

The sensors (Simplera, Guardian) are PMA approvals, while the pumps and the SmartGuard technology are 510(k) clearances, so the FDA categories differ. The company describes Abbott's Instinct as an iCGM made by Abbott, but its FDA category and number were not confirmed here Not yet confirmed. In Europe the company says the MiniMed 780G (with Instinct) received a CE mark in February 2026 and MiniMed Flex by September 2026. The status of approvals or certifications in Japan was not checked for this article.

A materials engineer's view: what happens when the enzyme layer's thickness leaves its specification

According to FDA's recall record, the Guardian 4 sensor recall that started on 15 November 2023 was due to the possibility that the thickness of the glucose oxidase (GOX) layer was outside its specification (5.5 to 9.2; the record gives the unit as "m", probably micrometres but not confirmed). A layer that is too thin could affect sensor values or cause loss of sensor function, and the record says the manufacturing team found it with Filmetrics (thin-film thickness) measurement Sourced. The quantity field shows 114,758 units and the root-cause category is "process change control". The enzyme layer is where glucose is converted into current, so its thickness is a design value that governs sensitivity and life. The record shows that the enzyme-layer thickness is measured optically during manufacturing as a control Not yet confirmed.

6. What it has done so far

WhenEventMeaning
28 September 2016The MiniMed 670G system received PMA approval (P160017). A system including technology (SmartGuard) that automatically adjusts basal insulinThe starting point for automated insulin delivery systems
8 March 2018The Guardian Connect system received PMA approval (P160007), for continuous or periodic monitoring of glucose in interstitial fluidA CGM approved independently of a pump
21 April 2023A change to the MiniMed 780G system, expanding it to include the Guardian 4 sensor, approved as a PMA supplementThe second-generation AID system
15 November 2023Recall of Guardian 4 sensors started; the GOX layer thickness may be out of specification (FDA record)114,758 units. See the materials view above
24 July 2024Simplera, a new single-use CGM, approved as a PMA supplement (P160007/S047)The company's own new-generation sensor
31 July 2024Global integration, supply and distribution agreement signed with Abbott (amended 1 June 2026)Lets Abbott-made CGM sensors be used in MiniMed's systems
7 May 2025Recall of a lot of Simplera Sensors started: the cap label may not adhere, compromising sterility (FDA record, 1,440 units)A quality issue in sensor manufacturing
6 August 2025Simplera Sync and Guardian 4 sensors received PMA approval (P250012)Simplera Sync launched in the U.S. in September 2025 (10-K)
December 2025Instinct sensor, made by Abbott, launched in the U.S. (10-K). The S-1 for the IPO was filed on 19 DecemberSensor choices rise to three
9 March 2026IPO completed: 28 million shares at US$20.00. Medtronic holds 252,813,348 shares (about 90%)An independent listed company, separated from Medtronic's diabetes business
12 March 2026MiniMed Flex pump received 510(k) clearance (K253743); U.S. launch in JuneA smaller new-generation pump
27 May 2026A software anomaly when the MiniMed 780G is paired with the Instinct sensor can make the pump stop delivering insulin under certain conditions; a software update (6.63) was issued (FDA record, 778 units)A system-integration issue. Facts only
1 June 2026Abbott agreement amended: adds dual glucose-ketone sensors designed to integrate exclusively with MiniMed's systems. Initial term of seven years from 1 June 2026As stated in the 10-K
14 September 2026Medtronic launched the exchange offer, letting Medtronic shareholders swap Medtronic shares for the MiniMed shares it holdsSee "What comes next" below

7. What comes next

EXCHANGE OFFER EXPIRY (EXPECTED)9 October 2026Unless extended or terminated. The final exchange ratio is to be announced by 9:00 a.m. New York time on the preceding trading day
FY2027 OUTLOOK (TO 30 APRIL 2027)About 10.5% organic revenue growthAdjusted EBITDA margin of about 16%. Growth outlook raised on 1 September 2026
MINIMED FIT (PATCH PUMP)U.S. launch expected summer 2027Submitted to FDA in September 2026
VIVERA (FULLY CLOSED LOOP)U.S. launch expected second half of 2027Enrollment in the U.S. pivotal trial for type 1 and type 2 complete

The terms of the exchange offer (Medtronic's announcement) are as follows Not yet confirmed.

  • Medtronic offers up to 225,361,295 MiniMed shares in exchange for Medtronic shares. One condition for completion is that at least 112,680,647 MiniMed shares be exchanged. The aim is to split off at least 80.1% of MiniMed shares.
  • The exchange is made at a 7% discount for Medtronic shareholders (up to 4.5939 MiniMed shares per Medtronic share; if the cap does not apply, about US$107.53 of MiniMed stock for every US$100 of Medtronic stock). Participation is voluntary.
  • If the offer is undersubscribed, Medtronic says it intends to dispose of its remaining shares through a spin-off, split-off, debt-for-equity exchange or a combination.
  • As of 4 October 2026, no result of the exchange offer appeared in the sources checked.

Product plans (results release of 1 September 2026; company outlook, not results): the next-generation extended-wear sensor has received a U.S. IDE (investigational device exemption) with the first pivotal-trial enrollment expected this fall. MiniMed Flex has received a CE mark with a European launch expected in November 2026.

What could not be confirmed

The final result and exchange ratio of the exchange offer / sales, profit and R&D for sensors alone / production capacity and investment for sensors other than Simplera / FDA category and number for Instinct (made by Abbott) / costs under the transition services agreements with Medtronic / the status of approvals or certifications in Japan / public grant funding. None appears in the primary sources checked for this article.

8. What the company is aiming for in biosensors

This section separates what the company says it is aiming for from what this article reads into the record. An Inference is our interpretation of primary sources, not a statement by the company.

What can be readSupporting primary informationCategory
Offering a choice of sensors to widen the pump's appealThe 10-K says its systems work with three sensors: Simplera Sync, Instinct and Guardian. The first-quarter results release credits new sensor launches for growth in new U.S. pump salesSourced / Inference (the intent)
Locking in the relationship with Abbott over the long termThe June 2026 amendment sets an initial seven-year term and adds an exclusive integration of Abbott-made glucose-ketone dual sensorsSourced
Investing in its own next-generation sensor (extended wear)U.S. IDE approved; first pivotal-trial enrollment expected in fall 2026 (company plan)Sourced (plan) / Inference (priority)
Raising the degree of automation of dosingEnrollment complete in the pivotal trial of the Vivera (fully closed-loop) algorithm; U.S. launch expected in the second half of 2027Sourced (company plan)
Raising capital and making decisions as an independent businessThe IPO, the start of the exchange offer, and Medtronic's statement about the "focus and flexibility" of an independent companySourced (company statement)
One inference in detail

The reading that the company is "offering a choice of sensors" rests on the fact that it has prepared regulatory documents so that its own and Abbott's sensors can be used with the same pump. The FDA summary for SmartGuard technology (K253585) records integration verification with a compatible iCGM, "Instinct, made by Abbott" Sourced. On the other hand, the May 2026 recall record describes a software anomaly that arises only with the Instinct pairing, which suggests that supporting several sensors also adds a burden of verifying each combination Not yet confirmed.

9. Map of partnerships in biosensors

MiniMed's collaborations, funding and agreements Only relationships stated in filings, company announcements or ClinicalTrials.gov; no press-based links Supply / integration / research Sales Equity / development funding MiniMed Pump + CGM Medtronic plc (parent) Holds about 90% (252,813,348 sh.) Exchange offer began 14 Sep 2026 Abbott Sensor supply and integration Amended June 2026; 7-year term Blackstone Development funding for Flex and Fit, in return for royalties Sole-source suppliers The 10-K cites reliance on sole suppliers as a risk; none named Clinical study partners 79 studies on ClinicalTrials.gov (sponsor or collaborator) DME suppliers, pharmacies, tenders Mainly the DME channel in the U.S.; national tenders elsewhere (10-K) Suppliers of materials and parts are not named
Fig. 2 MiniMed's collaborations, funding and agreements. Sources: MiniMed's Form 10-K (FY2026), Medtronic's exchange-offer announcement of 14 September 2026, and ClinicalTrials.gov. Line style shows the type of relationship (navy = supply, integration or research; orange = sales; dotted = equity or development funding; gray = counterparties not named). Relationships whose counterparties are not named in the sources are shown without names.

The 10-K names Abbott (supply and integration), Blackstone (co-development funding) and Medtronic (parent and separation agreement). The 10-K also says the Abbott agreement includes a commitment by MiniMed not to purchase, market or distribute CGMs from certain of Abbott's CGM competitors. Among the conditions under which Abbott may terminate, the 10-K lists a failure by MiniMed to buy Abbott CGMs for a continuous period, and a failure to launch certain Abbott CGMs in certain markets within four years Sourced.

10. Technologies, products and services

MiniMed's products combine an insulin pump (or smart pen), a CGM sensor, an algorithm that adjusts dosing automatically (SmartGuard), a smartphone app and a clinician-facing cloud (CareLink) Sourced. Much of the performance language comes from the company itself ("superior clinical results"), so this article treats it as company statements and does not rank values across products.

ProductWhat the official documents sayNote
MiniMed 780G system (AID)10-K: the SmartGuard algorithm adjusts pump dosing automatically every five minutes based on CGM readings; second-generation AID. FDA document (K253470): the 780G pump is for ages 7 and older and requires a prescriptionThe older 770G, 740G, 720G and 630G are also listed
MiniMed FlexFDA document (K253743): an ACE pump that communicates with compatible iCGMs, Medtronic CGMs and automated dosing software. Cleared March 2026, U.S. launch in June. Launched with Simplera and with Instinct (Q1 results release)A product that received Blackstone development funding
MiniMed Fit (patch pump)Submitted to FDA in September 2026; launch expected summer 2027 (Q1 results release)Not yet cleared; a company plan
MiniMed Go (smart MDI)10-K: a system for injection users that combines a smart insulin pen (InPen), a CGM sensor and an app, without a pumpFor users who prefer injections
Simplera / Simplera SyncFDA (PMA): the Simplera system was approved as a supplement on 24 July 2024; Simplera Sync received PMA approval on 6 August 2025 (P250012), intended to function as a component of an automated insulin dosing systemA single-use sensor of its own. The 10-K cites production challenges
Guardian 4FDA (PMA): added to the 780G system's scope on 21 April 2023 and also approved in P250012. A recall about enzyme-layer (GOX) thickness control in November 2023The previous-generation own sensor
Instinct (made by Abbott)10-K: a CGM sensor manufactured by Abbott Diabetes Care, launched in the U.S. in December 2025. FDA document (K253585) records integration testing with Instinct (made by Abbott) as an iCGMFDA category and number were not confirmed here
A materials engineer's view: with two kinds of sensor on one pump, differences in materials and processes become a combination-verification burden

The Guardian 4 recall record shows that keeping the enzyme (GOX) layer's thickness within specification bears directly on a sensor's values and life; it was found by Filmetrics thin-film measurement Sourced. The May 2025 recall of Simplera Sensors concerned a packaging and sterile-barrier problem: the cap label might not adhere and compromise sterility (1,440 units). In terms of materials, then, separate controls, over the enzyme layer (thickness) and over sterile packaging (label and cap adhesion), support the quality of the same sensor Not yet confirmed.

Moreover, MiniMed uses its own sensors (Simplera, Guardian) and Abbott's (Instinct) on the same pump. Because each sensor differs in electrode, membrane, adhesive and sterilization design, the pump's algorithm has to be verified to treat the outputs of sensors with different materials and characteristics on a common basis. That is probably why the FDA summary records "integration with Instinct" separately (Inference). The 10-K says clinical studies show "clinically equivalent glycemic outcomes" across Simplera Sync, Instinct and Guardian, but that is a company statement that this article has not verified.

11. The risks this company carries

RiskDetailCategory
Uncertainty from the separation and listingThe exchange offer is due to expire on 9 October 2026 (unless extended or terminated). If conditions such as the exchange of at least 112,680,647 shares are not met, Medtronic says it will dispose of the rest another way. The 10-K lists, as risks, a short record as a listed company, substantial costs of the separation and rebranding, and the chance that the expected benefits of the separation are not realizedSourced / outcome Not yet confirmed
Dependence on AbbottAbbott is the exclusive third-party CGM supplier for certain AID and smart MDI systems. Abbott may terminate if MiniMed fails to purchase for a set period, among other reasons. MiniMed has also agreed not to buy, market or distribute CGMs from certain Abbott competitorsSourced (10-K)
Supply and manufacturing of its own sensorsThe 10-K says the company has experienced reductions or interruptions in supply and manufacturing difficulties, including for the Simplera CGM. It has two manufacturing sites and also lists reliance on sole suppliers as a riskSourced (company's own risk statement)
Recalls and correction noticesFDA's recall database lists, for sensors, Guardian 4 (started 15 November 2023, 114,758 units), Simplera Sensor (started 7 May 2025, 1,440 units) and the pump-with-Instinct pairing (started 27 May 2026, 778 units), among others. There are also many correction notices for pumps and apps. Responses for individual users are set out in each notice; this article gives no advice to usersSourced (as recorded by FDA)
LossesFY2026 operating loss of US$190 million and net loss of US$317 million. A one-time charge of US$157 million in the fourth quarter of FY2026 for minimum royalties under the Blackstone agreement. Operating income of US$5 million in the first quarter of FY2027Sourced
Competition and reimbursementDexcom's 10-K names MiniMed (Medtronic's diabetes business) as a competitor (another company's filing). MiniMed's 10-K cites coverage and reimbursement terms as a riskSourced
Comparing fiscal yearsThe first quarter of FY2027 includes about one extra week of sales because of the 53-week year. Read growth rates with careSourced (company statement)
Verifying the sensor business from outsideSales, profit and R&D for sensors alone are not disclosed; only system-wide figures areNot yet confirmed

12. Glossary

AID (automated insulin delivery)
A system in which the pump automatically adjusts insulin dosing according to CGM readings.
ACE pump
An insulin pump designed to act on commands from external automated dosing software (an FDA category, 21 CFR 880.5730).
iCGM
An FDA category of CGM, defined by special controls, designed to connect safely to other devices.
PMA approval
FDA's premarket approval required for high-risk devices. It is a different category from 510(k) clearance.
510(k) clearance
FDA's finding that a device is substantially equivalent to a legally marketed predicate device.
GOX (glucose oxidase) layer
The enzyme layer formed on a sensor's electrode. Its thickness is a design value that bears on sensitivity and life.
Exchange offer
An offer allowing a parent company's shareholders to swap parent shares for shares of a subsidiary that the parent holds. Participation is voluntary.
Split-off / spin-off
Ways of separating a subsidiary: shareholders swap parent shares for subsidiary shares (split-off), or receive subsidiary shares free of charge (spin-off).
DME (durable medical equipment)
In the U.S., a specialist distribution channel for devices such as pumps and CGMs.
IDE (investigational device exemption)
FDA's approval to run a clinical study of an investigational device in the U.S.
Concept image: a sensor on the upper arm sends readings to a small pump at the waist and a smartphone app, which adjust insulin dosing
Fig. 3 Concept image of how an automated insulin delivery system works (AI-generated concept; it does not depict a real product or person). A small sensor worn on the arm sends readings to a small pump on a belt and to a smartphone, and the pump adjusts dosing.

13. References

  1. MiniMed Group Form 10-K for the fiscal year ended 24 April 2026 (filed 29 June 2026). Business description, sales and costs, employees, manufacturing sites, the Abbott and Blackstone agreements, risks. https://www.sec.gov/Archives/edgar/data/2062583/000162828026045915/mdt-20260424.htm
  2. MiniMed Group Form 10-Q for the quarter ended 31 July 2026 (filed 8 September 2026). Results for May to July 2026. https://www.sec.gov/Archives/edgar/data/2062583/000162828026060926/mdt-20260731.htm
  3. MiniMed Group Form 8-K (9 March 2026). Completion of the IPO (28 million shares at US$20.00) and the separation. https://www.sec.gov/Archives/edgar/data/2062583/000162828026016018/mmed-20260305.htm
  4. MiniMed Group Form 8-K Exhibit 99.1 (1 September 2026, first-quarter FY2027 results release). Results, FY2027 outlook and product plans. https://www.sec.gov/Archives/edgar/data/2062583/000162828026059699/exhibit991-fy27q1earningsr.htm
  5. Medtronic Press release filed under Form 425, "Medtronic Launches Exchange Offer to Complete Separation of MiniMed Group, Inc." (14 September 2026). Terms and timetable of the exchange offer. https://www.sec.gov/Archives/edgar/data/1613103/000162828026061674/a425-pressreleasexmdtexcha.htm
  6. Medtronic Form 8-K (17 September 2026). Commencement of the exchange offer and a correction to the prospectus. https://www.sec.gov/Archives/edgar/data/1613103/000162828026062505/mdt-20260917.htm
  7. Medtronic Form 10-K for the fiscal year ended 24 April 2026 (filed 18 June 2026). Group sales and Diabetes segment sales. https://www.sec.gov/Archives/edgar/data/1613103/000162828026044354/mdt-20260424.htm
  8. U.S. Food and Drug Administration (openFDA) PMA P250012 (Simplera Sync and Guardian 4 sensors; approved 6 August 2025). https://api.fda.gov/device/pma.json?search=pma_number:P250012
  9. U.S. Food and Drug Administration (openFDA) PMA P160017 (MiniMed 670G / 780G systems; approved 28 September 2016, with S091 on 21 April 2023). https://api.fda.gov/device/pma.json?search=pma_number:P160017&limit=200
  10. U.S. Food and Drug Administration (openFDA) PMA P160007 (Guardian Connect, approved 8 March 2018; Simplera system S047 on 24 July 2024). https://api.fda.gov/device/pma.json?search=pma_number:P160007&limit=100
  11. U.S. Food and Drug Administration 510(k) clearance letter and summary K253585 (SmartGuard technology, 14 January 2026). https://www.accessdata.fda.gov/cdrh_docs/pdf25/K253585.pdf
  12. U.S. Food and Drug Administration 510(k) clearance letter and summary K253470 (MiniMed 780G insulin pump, 15 January 2026). https://www.accessdata.fda.gov/cdrh_docs/pdf25/K253470.pdf
  13. U.S. Food and Drug Administration 510(k) clearance letter and summary K253743 (MiniMed Flex pump, 12 March 2026). https://www.accessdata.fda.gov/cdrh_docs/pdf25/K253743.pdf
  14. U.S. Food and Drug Administration (openFDA) Recall record, event 93675 (Guardian 4 sensors; started 15 November 2023). https://api.fda.gov/device/recall.json?search=res_event_number:93675
  15. U.S. Food and Drug Administration (openFDA) Recall record, event 96745 (Simplera Sensor; started 7 May 2025). https://api.fda.gov/device/recall.json?search=res_event_number:96745
  16. U.S. Food and Drug Administration (openFDA) Recall record, event 99207 (MiniMed 780G paired with Instinct; started 27 May 2026). https://api.fda.gov/device/recall.json?search=res_event_number:99207
  17. ClinicalTrials.gov Count of studies registered with "Medtronic MiniMed" as sponsor or collaborator (API; total of 79 confirmed). https://clinicaltrials.gov/api/v2/studies?query.spons=Medtronic%20MiniMed&countTotal=true&pageSize=1&fields=NCTId
  18. Abbott Press release (25 August 2026). The exclusive integration of Abbott-made glucose-ketone sensors with MiniMed's dosing systems. https://abbott.mediaroom.com/2026-08-25-Abbott-receives-FDA-authorization-for-worlds-first-dual-glucose-ketone-sensing-technology-for-people-with-diabetes
  19. Dexcom Form 10-K for the year ended 31 December 2025. The competition discussion (MiniMed as the spin-off of Medtronic's diabetes business). https://www.sec.gov/Archives/edgar/data/1093557/000109355726000027/dxcm-20251231.htm
  20. MiniMed Official website (source of the logo image). https://www.minimed.com/

14. Claim-to-source audit

Claim in the articleCategorySource
Legal name MiniMed Group, Inc., formerly Medtronic's Diabetes business, Nasdaq listing (MMED), headquarters in Northridge, two manufacturing sites (California and Puerto Rico), about 8,000 employees, more than 2,000 patentsSourcedReference 1 https://www.sec.gov/Archives/edgar/data/2062583/000162828026045915/mdt-20260424.htm
FY2025 and FY2026 sales (US$2,715 / 3,102 million), gross profit, R&D, operating result, net result, capital expenditure; about 82% recurring revenue; about 70% outside the U.S.SourcedReference 1 https://www.sec.gov/Archives/edgar/data/2062583/000162828026045915/mdt-20260424.htm
Medtronic plc group sales (US$33,537 / 36,364 million) and the Diabetes segment (US$2,755 / 3,112 million)SourcedReference 7 https://www.sec.gov/Archives/edgar/data/1613103/000162828026044354/mdt-20260424.htm
Diabetes segment at about 8.6% of group sales, and the US$10 million gap between MiniMed's and Medtronic's sales figuresOur calculationOur own division and subtraction of figures in References 1 and 7. Reference 1 https://www.sec.gov/Archives/edgar/data/2062583/000162828026045915/mdt-20260424.htm
May to July 2026 sales of US$843 million (+16.6%), U.S. and international split, operating income, R&D, new pump sales, the 53-week effectSourcedReference 2 https://www.sec.gov/Archives/edgar/data/2062583/000162828026060926/mdt-20260731.htm Reference 4 https://www.sec.gov/Archives/edgar/data/2062583/000162828026059699/exhibit991-fy27q1earningsr.htm
IPO (9 March 2026; 28 million shares at US$20.00), the separation and Medtronic's 252,813,348 shares (about 90%)SourcedReference 3 https://www.sec.gov/Archives/edgar/data/2062583/000162828026016018/mmed-20260305.htm Reference 5 https://www.sec.gov/Archives/edgar/data/1613103/000162828026061674/a425-pressreleasexmdtexcha.htm
Cash from the share issuance of US$538 millionSourcedReference 1 https://www.sec.gov/Archives/edgar/data/2062583/000162828026045915/mdt-20260424.htm
Blackstone co-development agreements (Flex and Fit; US$157 million one-time charge; termination payments up to US$216 million); assigned in the separationSourcedReference 1 https://www.sec.gov/Archives/edgar/data/2062583/000162828026045915/mdt-20260424.htm
Abbott agreement (signed 31 July 2024, amended 1 June 2026, initial seven-year term, exclusive third-party CGM supplier, commitment not to buy certain competitors' CGMs, termination conditions)SourcedReference 1 https://www.sec.gov/Archives/edgar/data/2062583/000162828026045915/mdt-20260424.htm
Planned exclusive integration of Abbott-made glucose-ketone sensors with MiniMed systems (expected in 2027)Not yet confirmedReference 18 https://abbott.mediaroom.com/2026-08-25-Abbott-receives-FDA-authorization-for-worlds-first-dual-glucose-ketone-sensing-technology-for-people-with-diabetes
Registered studies on ClinicalTrials.gov (79)SourcedReference 17 https://clinicaltrials.gov/api/v2/studies?query.spons=Medtronic%20MiniMed&countTotal=true&pageSize=1&fields=NCTId
Sensor-only splits of R&D and capital expenditure, sensor acquisitions or joint ventures, NIH or other public funding, FDA category of Instinct, the result of the exchange offerNot yet confirmedNot found in References 1, 2 and 5. We make no estimate. Reference 1 https://www.sec.gov/Archives/edgar/data/2062583/000162828026045915/mdt-20260424.htm
MiniMed 780G: PMA supplement P160017/S091 (21 April 2023); original MiniMed 670G approval (28 September 2016)SourcedReference 9 https://api.fda.gov/device/pma.json?search=pma_number:P160017&limit=200
Simplera system: PMA supplement P160007/S047 (24 July 2024); Guardian Connect (8 March 2018)SourcedReference 10 https://api.fda.gov/device/pma.json?search=pma_number:P160007&limit=100
Simplera Sync and Guardian 4 sensors: PMA approval P250012 (6 August 2025, product code SFI)SourcedReference 8 https://api.fda.gov/device/pma.json?search=pma_number:P250012
SmartGuard technology: 510(k) K253585 (14 January 2026, product code QJI), ages, integration testing with Instinct (made by Abbott)SourcedReference 11 https://www.accessdata.fda.gov/cdrh_docs/pdf25/K253585.pdf
MiniMed 780G insulin pump: 510(k) K253470 (15 January 2026, QFG, ages 7 and older, prescription required)SourcedReference 12 https://www.accessdata.fda.gov/cdrh_docs/pdf25/K253470.pdf
MiniMed Flex pump: 510(k) K253743 (12 March 2026, QFG, ACE pump)SourcedReference 13 https://www.accessdata.fda.gov/cdrh_docs/pdf25/K253743.pdf
U.S. launches (Simplera Sync in September 2025, Instinct in December 2025, Flex in June 2026) and European CE marks (780G with Instinct in February 2026, Flex)SourcedReference 1 https://www.sec.gov/Archives/edgar/data/2062583/000162828026045915/mdt-20260424.htm Reference 4 https://www.sec.gov/Archives/edgar/data/2062583/000162828026059699/exhibit991-fy27q1earningsr.htm
Guardian 4 recall: started 15 November 2023, GOX layer thickness possibly out of specification, found by Filmetrics measurement, 114,758 unitsSourcedReference 14 https://api.fda.gov/device/recall.json?search=res_event_number:93675
Simplera Sensor recall: started 7 May 2025, cap label may not adhere and compromise sterility, 1,440 unitsSourcedReference 15 https://api.fda.gov/device/recall.json?search=res_event_number:96745
Recall for the 780G paired with Instinct: started 27 May 2026, pump stops dosing, update 6.63, 778 unitsSourcedReference 16 https://api.fda.gov/device/recall.json?search=res_event_number:99207
Terms of the exchange offer (up to 225,361,295 shares, at least 112,680,647, 7% discount, cap of 4.5939 shares, expected expiry 9 October 2026, disposal of the remainder if undersubscribed)Not yet confirmedReference 5 https://www.sec.gov/Archives/edgar/data/1613103/000162828026061674/a425-pressreleasexmdtexcha.htm Reference 6 https://www.sec.gov/Archives/edgar/data/1613103/000162828026062505/mdt-20260917.htm
FY2027 outlook (about 10.5% organic growth, adjusted EBITDA margin of about 16%) and plans for MiniMed Fit, Vivera, the next-generation sensor and the Flex launch in EuropeNot yet confirmedReference 4 https://www.sec.gov/Archives/edgar/data/2062583/000162828026059699/exhibit991-fy27q1earningsr.htm
The 10-K's risk statements (Simplera CGM supply interruptions, sole suppliers, costs of separation and rebranding)SourcedReference 1 https://www.sec.gov/Archives/edgar/data/2062583/000162828026045915/mdt-20260424.htm
Dexcom's 10-K describing MiniMed as the spin-off of Medtronic's diabetes business and as a competitorSourcedReference 19 https://www.sec.gov/Archives/edgar/data/1093557/000109355726000027/dxcm-20251231.htm
The four-type classification (diabetes and continuous monitoring primary) and example company namesOur calculationOur own classification. Company names are examples from this series. Reference 1 https://www.sec.gov/Archives/edgar/data/2062583/000162828026045915/mdt-20260424.htm
Readings of a choice of sensors, locking in the Abbott relationship and investing in a next-generation sensorInferenceOur interpretation of References 1, 4, 11 and 18. Reference 1 https://www.sec.gov/Archives/edgar/data/2062583/000162828026045915/mdt-20260424.htm
Readings of enzyme-layer and sterile-packaging controls underpinning quality, and of the burden of verifying different sensorsInferenceOur interpretation of References 14, 15, 16 and 11. Reference 14 https://api.fda.gov/device/recall.json?search=res_event_number:93675

Last updated 4 October 2026 / Troy Technical
Every figure in this article comes from MiniMed Group's statutory filings (SEC filings), U.S. FDA public documents, ClinicalTrials.gov and the company's official announcements. No research-firm estimates or press-based figures have been used. Passages marked "Inference" are this article's interpretation of primary sources, not statements made by the company. Nothing here recommends using a product or makes a medical judgment.