TECHNOLOGY EXPLAINER
Electrochemical Sensors
— Reading Current and Potential, and the Materials Behind Electrodes and Membranes
From blood glucose test strips to sweat electrolyte sensors, most biosensors work by measuring electrons moving in and out at an electrode surface. This article is written for materials engineers: it sets out what is measured, what the electrodes and membranes are made of, and where things get difficult. It is not medical advice.
- What an electrochemical sensor is, in three lines
- Three things to measure: current, potential and impedance
- Why three electrodes?
- Our calculation: current depends on when you read it, and potential acts on a log scale
- Three generations of enzyme electrodes: how electrons are handed over
- Electrode materials: carbon, gold, platinum and silver/silver chloride
- A materials engineer's view (1): the ink decides the electrochemistry
- How screen-printed electrodes are made, and their layer stack
- What membranes do: diffusion limiting, exclusion and biocompatibility
- A materials engineer's view (2): a membrane is a trade-off between what it lets through and what it stops
- Interference and selectivity: how they are tested
- Strengths, challenges and what we could not confirm
- Glossary / References / Claim-to-source audit
Sourced = stated in published material (source linked)
Our calculation = a value this article derived from explicitly stated assumptions
Not yet confirmed = a plan or target with no confirmed outcome
Materials-design readings are marked Commentary; weakly grounded guesses are marked Inference.
Performance figures are given together with what they were compared against, what sample was used and under what conditions. Please do not read in vitro results as clinical performance in everyday life.
1. What an electrochemical sensor is, in three lines
An electrochemical sensor is "a sensor that reads the oxidation and reduction (the transfer of electrons) occurring at an electrode surface as a current or a potential". Combine it with a biological recognition element (an enzyme, antibody or nucleic acid) and you have an electrochemical biosensor.
- Definition: IUPAC's recommendation defines an electrochemical biosensor as a self-contained, integrated device that gives specific quantitative or semi-quantitative analytical information using a biological recognition element held in direct spatial contact with an electrochemical transduction elementSourced. A device that needs extra steps such as adding reagents should be called a "bioanalytical system", and a device that is single-use and cannot monitor continuously should be called a "single-use biosensor"Sourced
- Where it began: Clark's oxygen electrode of 1955–56 is the starting point; a 1962 report described the first glucose sensor combining an enzyme with an electrode, and the first dedicated commercial glucose biosensor followed in 1975 (Yellow Springs Instrument)Sourced (the review by Grieshaber et al.)
- Seen as a materials problem: the signal comes from a stack of layers: the electrode material, the membranes over it, the way the enzyme and mediator are held in place, and the substrate and sealing. Most of the difficulty shows up as materials and manufacturing issues: interfering substances, protein fouling of membranes, temperature, storage stability and unit-to-unit variation
This article explains materials, principles, evaluation and regulation. It gives no advice to individual readers about diagnosis, treatment or how to read a measured value. It does not rank any particular product or company.
2. Three things to measure: current, potential and impedance
How an electrochemical sensor reads out falls broadly into three groups. The review by Grieshaber et al. divides them into reactions that generate a current (amperometry), reactions that generate a potential or charge accumulation (potentiometry), and reactions that change the conductivity of the medium (conductometry), and lists impedance measurement and field-effect transistors (FETs) alongside themSourced.
| Method | What is measured | Typical use | Materials-side watch points |
|---|---|---|---|
| Current measurement (amperometry / voltammetry) | The electrode potential is held constant (or swept) and the current from oxidation or reduction is measured | Oxidase-enzyme sensors such as glucose sensors; mediator-assisted current readout | Electrode area and surface state, diffusion resistance of the membrane, and interferents that oxidise at the same potential |
| Potential measurement (potentiometry) | The potential difference between working and reference electrodes is measured with almost no current flowing | Ion-selective electrodes (pH, Na⁺, K⁺ and others) | Reference electrode stability; composition and lifetime of the ion-selective membrane (PVC, ionophore and so on) |
| Impedance measurement | A small alternating current is applied and the response (resistive and capacitive parts) is measured at each frequency | Detecting interface changes from binding of antibodies and the like; skin and tissue measurements | Hard to interpret. Covered in detail in our explainer on impedance measurement in this series |
| Field-effect transistor | A potential change at the gate surface is read as a transistor current | ISFETs, nanowire and graphene FETs | Debye screening and reproducibility. Covered in detail in our explainer on FET sensors in this series |
All Sourced (Grieshaber et al. 2008 [Ref. 2]). The "Materials-side watch points" column is this article's commentary.
This article concentrates on current and potential measurement and digs into electrode and membrane materials. The idea of reading binding without labels is covered in our explainer on label-free detection and the limit of detection in this series.
3. Why three electrodes?
The basic electrochemical sensor has three electrodes: working, reference and counter. The review by Grieshaber et al. explains that the reference electrode is kept away from the reaction site to hold a known, stable potential (usually Ag/AgCl), the working electrode acts as the transducer for the biochemical reaction, and the counter electrode makes the connection to the electrolyte so that current can be passed through the working electrodeSourced. These electrodes must be conductive and chemically stable, and platinum, gold, carbon (graphite and similar) and silicon compounds are used depending on the targetSourced.
If current flows through the reference electrode its potential drifts, and it becomes unclear what the set potential is measured againstOur calculation. For example, with a sample resistance of 1 kΩ and a current of 10 µA, Ohm's law gives a potential drop of 1,000 Ω × 10 µA = 10 mV. An error of 10 mV is not negligible when the potential has to be set precisely (commentary). With three electrodes the current flows only between the working and counter electrodes and the reference electrode just senses potential, which avoids this error.
Assumptions and limits: 1 kΩ and 10 µA are illustrative values chosen for this article; real values vary widely with electrode size, sample and layout. Depending on electrode layout and sample the effect can be small, so it does not dominate in every sensor.
The reference electrode is the plainest part, and the easiest to break
The Ag/AgCl reference electrode is one of the simplest and most practical reference electrodes. Its standard potential is reported as about 0.22 V at unit chloride activitySourced (Shinwari et al.). Miniaturising it causes trouble. When AgCl dissolves, the ion activity in the electrolyte changes and the potential drifts, and the smaller the volume of solution, the worse this becomes. Thin layers are used up quickly, exposing the metal underneath and producing mixed potentialsSourced. Microfabricated versions generally use some kind of diffusion barrier to limit dissolutionSourced.
Where Ag/AgCl paste is applied directly by screen printing, adhesion to the support may not be good, and reliability depends on the homogeneity of the paste, as Shinwari et al. point outSourced. An Ag/AgCl electrode without an internal filling solution is used as a "pseudo-reference electrode", and its potential is constant only in a particular sample that does not change over timeSourced. That premise breaks down for samples such as sweat, where chloride concentration can vary Inference (a general inference from Shinwari et al.; we could not confirm validation for individual products).
4. Our calculation: current depends on when you read it, and potential acts on a log scale
Current measurement and potential measurement give signals that behave quite differently. The review by Grieshaber et al. explains that when the working electrode is held at a constant potential, the time course of the current follows the Cottrell equation, with the current depending on the number of electrons n, the Faraday constant F, the electrode area A, the concentration c₀, the diffusion coefficient D and time tSourced. In potentiometry, the relation between concentration and potential is governed by the Nernst equationSourced. The calculations below simply put illustrative numbers into those two equations.
- Current decay: putting the assumptions above into the Cottrell equation, i = nFAc₀√D / √(πt), gives about 10.2 µA at 1 s, about 3.2 µA at 10 s and about 1.9 µA at 30 sOur calculation. The current falls in inverse proportion to √t. Even for the same sample, the value depends on how many seconds after the start you read the current (commentary)
- Nernst slope: the ideal response for a monovalent ion is 59.2 mV per decade at 25 °C (RT/F × ln10), and about 29.6 mV per decade for a divalent ion. At 37 °C (310.15 K), a monovalent ion gives about 61.5 mV per decadeOur calculation. Going from 25 °C to 37 °C alone changes the slope by about 4%, so temperature compensation becomes part of the design in potentiometry (commentary)
- Upper bound on charge: 0.5 µL of a sample with 100 mg/dL glucose (about 5.55 mM) contains about 2.8 nmol of glucose. Oxidation of hydrogen peroxide is a two-electron processSourced (Weltin et al.), and glucose plus oxygen yields one hydrogen peroxide per glucoseSourced (Grieshaber et al.). Taking two electrons per glucose, the upper bound on charge if everything reacted is about 2 × 96,485 C/mol × 2.8 nmol ≈ about 540 µCOur calculation
Assumptions and limits: these are the equations for ideal conditions: a disc electrode, planar diffusion and a reversible one-electron reaction. Real strips add diffusion through membranes and reaction layers, enzyme kinetics and mediator behaviour, so the current's time course does not follow this equation exactly. The diffusion coefficient of 7×10⁻⁶ cm²/s is our assumption and varies with the molecule and solvent. The 0.5 µL and 100 mg/dL are also illustrative. Nothing here describes the reading time or sample volume of any individual product.
Potentiometry in practice: the ideal slope is not always reached
In potentiometry, ion-selective electrodes (ISEs) perform best at low concentrations. The review notes that detection limits of 10⁻⁸ to 10⁻¹¹ M have been reached in some cases, but that the range of ions that can be measured is limited, with nickel, manganese, mercury and arsenate ions among those lacking coverageSourced. For an FET-based platform that measures sodium, potassium and calcium in sweat, one group reported in vitro results of −57 mV per decade (Na⁺), −48 mV per decade (K⁺) and −26 mV per decade (Ca²⁺)Sourced (cited in the review by Nguyen et al.). Against the ideal values above (about 59 mV per decade for monovalent and about 30 mV per decade for divalent ions), Na⁺ is close while K⁺ and Ca²⁺ are lower. However, these come from different devices and measurement conditions placed side by side; this is not a ranking of methods.
5. Three generations of enzyme electrodes: how electrons are handed over
In electrochemical biosensors that use enzymes, "generations" are distinguished by how the electrons from the enzyme reaction reach the electrode. In the grouping by Grieshaber et al., the first generation uses oxygen as the electron acceptor and produces hydrogen peroxide; the second generation replaces oxygen with a mediator (such as an iron-based ferro/ferricyanide); and the third generation connects the enzyme directly to the electrode, with no mediatorSourced.
- Weakness of the first generation: oxidising hydrogen peroxide requires a relatively high potential, and at that potential reducing substances such as ascorbic acid, uric acid and acetaminophen are also oxidised and add to the current. There are two main remedies: (i) keep interferents away from the electrode with a selective membrane (Nafion or cellulose acetate), or (ii) lower the working potential into the 0–0.2 V (vs Ag/AgCl) region so that the interferents themselves do not reactSourced (Zhu et al.)
- Place of the second generation: mediators are artificial electron acceptors that carry electrons between enzyme and electrode; they are small, mobile molecules, and can also be immobilised together with the enzymeSourced (Grieshaber et al.). Examples given include ferrocene and its derivatives, ferrocyanide, methylene blue and benzoquinoneSourced
- Weakness of the third generation: direct electron transfer slows as the enzyme moves away from the electrode, and the signal has been reported to fall exponentially with distanceSourced (Grieshaber et al.). The orientation and distance of the enzyme must be controlled through materials (commentary)
Enzymes, mediators and electrode-interface materials are covered in detail in our explainer on enzymes and electrode interface materials in this series.
6. Electrode materials: carbon, gold, platinum and silver/silver chloride
Here we line up the main working-electrode materials and the standard reference electrode, looking at what each does well and what needs care. The table reorganises statements in published reviews from a materials viewpoint.
| Material | Main role | Strengths | Points of care |
|---|---|---|---|
| Carbon (graphite, carbon ink, glassy carbon) | Working / counter electrode | Low background current, wide potential window and chemically inertSourced. Can be made into printable ink | The ink is a mixture of particles, a polymer binder and additives, and its formulation is often a trade secret. Particle type, size and loading change electron transferSourced |
| Gold | Working electrode (thin film, printed) | Easy to make a smooth surface, and recognition elements can be attached through thiol self-assembled monolayers (SAMs)Sourced | A SAM must be designed so that it does not hinder electron or analyte diffusionSourced. Adhesion to the substrate can become an issue in thin films (commentary) |
| Platinum | Working electrode (oxidising hydrogen peroxide in first-generation sensors) | Hydrogen peroxide is oxidised efficiently at polarised noble-metal electrodes such as platinum (a two-electron process)Sourced | Interferents are also oxidised at the same potential, so it is often combined with a selective membraneSourced |
| Silver / silver chloride (Ag/AgCl) | Reference electrode | One of the simplest and most practical reference electrodes. Standard potential about 0.22 V at unit chloride activitySourced | AgCl dissolution shifts the potential, and the smaller the electrode, the shorter its life. For printed paste, adhesion and paste homogeneity govern reliabilitySourced |
Carbon: Hayat and Marty [Ref. 3] / gold and reference-electrode background: Grieshaber et al. [Ref. 2] / platinum: Weltin et al. [Ref. 5] / Ag/AgCl: Shinwari et al. [Ref. 4]. The "Strengths" and "Points of care" groupings are this article's commentary.
The thin-film route: evaporating platinum onto polyimide
Electrodes can also be made not by printing but by evaporation or sputtering plus lithography. The review of enzyme-based microsensors (Weltin et al.) lists constructions such as theseSourced.
- A sensor with two working electrodes and an Ag/AgCl reference, made by vapour-depositing platinum on a 100 µm polyimide film and insulating it with another 2 µm polyimide layer
- A sensor whose platinum electrodes on 50–100 µm polyimide film are insulated on both sides with a 5 µm SU-8 epoxy resist
- Platinum sputtered onto a ceramic substrate (insulated with 10 µm polyimide), and platinum evaporated onto silicon (insulated with a CVD nitride film)
In each case the construction is a set of three: electrode metal, substrate and insulating layer, and the choice of materials translates directly into differences in stiffness, adhesion and processing cost (commentary).
7. A materials engineer's view (1): the ink decides the electrochemistry
Screen-printing inks for electrodes consist of particles (carbon or metal), a polymer binder, and additives that adjust dispersion, printability and adhesion, and the exact formulations are not disclosed by the companies that hold them as patents or trade secretsSourced (Hayat and Marty). That review states that differences in ink composition, such as particle type, size and loading, change the electron-transfer process substantiallySourced. Even under the same name "carbon electrode", a different ink gives different electrochemical behaviour.
- The binder is often an insulator: the polymer filling the gaps between conductive particles may partly cover the particle surfaces, so the apparent area and the electrochemically active area can differ Inference (reasoned from general knowledge of conductive composites; we could not confirm measurements for specific inks)
- Surface treatment brings out activity: one study described pre-anodising a printed carbon electrode to create defect or edge-plane-like sites and oxygen functional groups; for uric acid and dopamine detection the overpotential fell and the oxidation peaks separatedSourced (Hayat and Marty; a comparative study with oxygen-plasma treatment). It is one way for materials to narrow lot-to-lot spread (commentary)
- Weakness of mixing the mediator into the ink: mixing the mediator into the paste has been used for a long time, but it is described as having the drawbacks that it is unstable and difficult to mass-produce reproduciblySourced. Modification with metal oxides is being explored as an alternativeSourced
In other words, part of the lot-to-lot spread in sensors arises before any electrochemistry, as variation in the ink's particles, binder and drying conditions (commentary). For the adhesion and sealing side, the articles in our adhesives and encapsulants series are useful.
8. How screen-printed electrodes are made, and their layer stack
According to the review by Hayat and Marty, a screen-printed electrode is three electrodes (working, reference and counter) screen-printed on a chemically inert substrate. The steps are: (1) an inert substrate, (2) printing the working and auxiliary (counter) electrodes, (3) printing the reference electrode, (4) printing a protective (insulating) paste, and (5) modifying the working electrode for the target analyteSourced. Grieshaber et al. also note a growing trend of making electrodes by printing, which is low-cost and easy to mass-produceSourced.

View the original SVG schematic, source notes and numerical details
| Fabrication route | Electrode metal / substrate / insulation (examples from the reviews) | Differences from a materials viewpoint |
|---|---|---|
| Thick film (screen printing) | Conductive inks (carbon, silver, Ag/AgCl and others) printed on an inert substrate and insulated with a protective pasteSourced | Low cost and suited to volume productionSourced. Ink composition and drying conditions are a source of spread (commentary) |
| Thin film (evaporation / sputtering plus lithography) | Platinum evaporated or sputtered on polyimide, ceramic or silicon and insulated with polyimide, SU-8 or a nitride filmSourced (Weltin et al.) | Electrode dimensions can be made small and precise. Equipment and process costs tend to be higher (commentary). The substrate is far stiffer than tissueSourced |
Thick film: Hayat and Marty [Ref. 3], Grieshaber et al. [Ref. 2] / thin film: Weltin et al. [Ref. 5]. The parts of the right-hand column marked (commentary) are this article's own grouping.

9. What membranes do: diffusion limiting, exclusion and biocompatibility
An enzyme electrode is a structure of several membranes stacked on the electrode. The review of enzyme-based microsensors (Weltin et al.) describes this structure as followsSourced.
- Enzyme membrane: the enzyme has to be immobilised in a membrane or matrix without losing function. The most common method is to cross-link the enzyme with bovine serum albumin (BSA) using glutaraldehyde; a droplet is applied and the cross-linking proceeds as it dries
- Diffusion-limiting / selective membrane: a membrane such as Nafion or polyurethane may be added on top of the enzyme membrane. Applying a thin polyurethane layer is a common way to widen the linear range
- Exclusion (permselective) membrane: to keep interferents from reaching the electrode, it stops them by size exclusion or charge exclusion. Electropolymerised aminobenzenes (poly-o-phenylenediamine, poly-m-phenylenediamine) and drop-cast Nafion are typical examples. Small neutral molecules such as hydrogen peroxide pass, while interferents of higher molecular weight or with a charge are stopped
The same review states plainly that "widening the linear range means limiting diffusion, which costs sensitivity" and that adding a selective membrane also lowers sensitivitySourced. Highly sensitive sensors have a linear range limited to a few hundred micromolar, the same order as the oxygen concentration availableSourced.
Degradation on or in the body: from protein adsorption to encapsulation
In uses that touch the body, the membrane surface changes as it meets the sample. Weltin et al. outline the course of events for sensors placed in vivo in these stagesSourced.
- Seconds to hours: molecules and proteins adsorb to the surface. A large loss of sensitivity has been reported, especially for glucose. The cause is still debated. Some of it is reversible
- Days: a foreign-body reaction develops, molecules and cells gather around the implant, and sensor function is further impaired
- Eventually: encapsulation occurs and sensor performance is severely affected (in many cases beyond the sensor's lifetime)
There are two broad countermeasures: passively coating the surface with various materials, and actively releasing substances from the sensor membrane to soften the foreign-body reaction (for example nitric oxide release)Sourced. The classic review of biofouling reduction (Wisniewski and Reichert) examines nine kinds of modification: hydrogels, phospholipid-based biomimicry, flow-based systems, Nafion, surfactants, naturally derived materials, covalent attachment, diamond-like carbon, and surface topologySourced. Hydrogels are covered in our hydrogel explainer in this series, and adhesion to skin in our explainer on patch sensors and skin adhesion.
Mechanical mismatch is also pointed out. Metal needles, silicon, ceramics and most polymers, including polyimide, have an elastic modulus several orders of magnitude higher than tissue (especially the central nervous system), and true long-term biocompatibility is said to call for matching flexibility and elasticitySourced (Weltin et al.).
10. A materials engineer's view (2): a membrane is a trade-off between what it lets through and what it stops
The membrane over the electrode is asked to do the following three things at once.
- Let glucose (or hydrogen peroxide) through but stop interferents (selectivity)
- Restrict glucose influx relative to oxygen so the response stays linear over a wide concentration range (diffusion limiting)
- Keep proteins from sticking (antifouling), and even once fouling progresses, keep the sensitivity from changing much
The design variables of a membrane (pore size, charge, hydrophilicity, thickness) compete across these three. Weltin et al. also state that a selective membrane lowers sensitivity and that extending the linear range sacrifices itSourced. So optimising a membrane is not a single answer but a search for a compromise for each use (sample, duration of use, required accuracy) (commentary).
- Limits of antifouling: it is safer to read a passive coating as something that delays or weakens adhesion, not as something that brings it to zero (commentary). Weltin et al. note that all the sensors in their comparison table use the passive approach, and that the cause of the sensitivity loss from adsorption is still debatedSourced
- Evaluation tip: selectivity is often expressed as "sensitivity to the analyte divided by sensitivity to the interferent", but for an electrode whose sensitivity to the interferent is inherently low this ratio comes out high. The practical advice is therefore to measure the bare electrode's sensitivity to the interferent first and compare it with the membrane-coated electrode, which evaluates the membrane's effect directlySourced (Weltin et al.)
11. Interference and selectivity: how they are tested
IUPAC's recommendation lists the performance criteria for biosensors as including sensitivity, operational and linear range, detection and quantification limits, selectivity, response times, sample throughput, reproducibility, stability and lifetimeSourced. In electrochemical sensors selectivity (interference) is an especially pressing issue. Yoo and Lee explain that an electrochemical interferent is a species that is electroactive at the operating potential of the amperometric sensor, and that by donating electrons not derived from glucose it pushes the apparent value up in bloodSourced.
Interferents are not only the "electroactive" ones
- Haematocrit: Yoo and Lee note that oxygen from red blood cells can compete with the mediator for electrons when GOx is used, and that higher blood viscosity slows diffusion of all components and lowers the current; they describe haematocrit as among the largest sources of error in strip-based glucose measurementSourced
- Enzyme selectivity: PQQ-dependent glucose dehydrogenase (GDH-PQQ) also oxidises sugars other than glucose (maltose, maltotriose, metabolites of icodextrin and others), and overestimation has been reported on GDH-based devices when icodextrin is used in peritoneal dialysisSourced (Yoo and Lee). This is an example where the selectivity of the recognition element is decided on the enzyme side, not the electrode side (commentary)
- Drugs: acetaminophen is described as being oxidised directly once it diffuses through a porous membrane to the electrode, producing an interfering current that raises the glucose readingSourced (Yoo and Lee)
How regulators have interference tested
In a September 2020 guidance (non-binding recommendations) for 510(k) submissions of over-the-counter self-monitoring blood glucose systems, the US FDA sets out how interference should be evaluatedSourced. The main points are as follows.
- Test blood is prepared at three levels, roughly 50–70, 110–130 and 225–270 mg/dL target glucoseSourced
- Each interferent is tested at the highest concentration that could be encountered, and if significant interference appears it is diluted stepwise to find the concentration at which interference beginsSourced
- Test samples with the interferent added and control samples with only the solvent are measured with at least three strip lots (30 replicates per sample in total across three lots), and the relative bias is submitted with 95% confidence intervalsSourced
- Conditions known to interfere, such as ketoacidosis, are to be included as limitations in the labellingSourced
| Interferent (examples) | Recommended test concentration |
|---|---|
| Acetaminophen | 20 mg/dL |
| Ascorbic acid | 6 mg/dL |
| Uric acid | 23.5 mg/dL |
| Dopamine | 0.09 mg/dL |
| Maltose | 480 mg/dL |
| Galactose | 60 mg/dL |
| Haemoglobin | 1000 mg/dL |
| Triglycerides | 1500 mg/dL |
All Sourced (excerpted from Table 3 of the FDA guidance [Ref. 9], which lists 26 items as "the minimal list of substances that should be tested for interference"). These concentrations are recommended test levels. They are not safety values and do not indicate that concentrations below them have no effect.
Yoo and Lee state that the FDA documents refer to ISO 15197 for evaluation of precision, accuracy and user performance, and to the CLSI guideline EP-7A for interference testingSourced. ISO 15197:2013 (second edition, published May 2013) is a standard that sets requirements, design-verification procedures and validation of performance by intended users for in vitro blood-glucose monitoring systems that measure glucose in capillary blood, aimed at systems for self-testing by lay persons at homeSourced (the ISO official page). It was reviewed and confirmed in 2018Sourced. The text of the standard is paid, and we could not confirm its acceptance criteria, so they are not given here.
12. Strengths, challenges and what we could not confirm
Strengths of electrochemical sensors (from a materials viewpoint)
- Electrical signals make miniaturisation and low cost easy: they can be mass-produced by printing or evaporation, and small electrodes connect easily to electronicsSourced (Grieshaber et al. describe printing as "low-cost and easy to mass-produce")
- High freedom in materials: electrodes (carbon, gold, platinum), membranes (Nafion, polyurethane, electropolymerised films) and substrates (polyimide, ceramic, silicon) can be recombined into stacks suited to the useSourced (Weltin et al. and others)
Challenges
- Miniaturising the reference electrode: Ag/AgCl loses lifetime as it shrinks because of dissolution, and for printed paste, adhesion and homogeneity govern reliabilitySourced
- Interference and selectivity: in current-based methods, any species that is electroactive at the working potential interferes. Membranes, potential, subtraction and enzymes are combined to suppress it, but each trades off against sensitivity or linear rangeSourced
- Membrane degradation: in uses that touch the body, protein adhesion, foreign-body reaction and encapsulation change performance over time. A passive coating is a way of delaying this, not of eliminating it (commentary)
- Inks and lot-to-lot spread: the exact ink formulation is not disclosed, and differences in particles and binder change electron transferSourced
What we could not confirm
- The full text of the IUPAC recommendation: IUPAC's official site could not be opened by automated retrieval, so this article uses only what can be confirmed from the paper's abstract in PubMed.
- The numerical acceptance criteria of ISO 15197:2013: the standard is paid, and only the scope and edition information could be confirmed on the official page.
- The current edition of the CLSI interference-testing document: we did not confirm the latest edition or title of the successor to the EP-7A cited by Yoo and Lee.
- Specific ink formulations and the layer structure of real products: not confirmable from public information. Fig. 4 is an example.
- Long-term performance in or on the body: this article explains the principles of sensor materials and does not evaluate the clinical performance of any product. The regulatory position of individual products is covered in our explainer on regulatory pathways for medical devices and home use in this series.
- An electrochemical sensor reads the electron transfer at an electrode surface as a current or a potential. IUPAC defines an electrochemical biosensor as a self-contained device in which the recognition and transduction elements are integratedSourced
- Three electrodes (working, reference and counter): current flows between working and counter, and the reference electrode serves as the potential referenceSourced
- Under ideal conditions the current falls in inverse proportion to √t, and the Nernst response of a monovalent ion is about 59 mV per decade at 25 °COur calculation
- The electrode materials are carbon, gold, platinum and Ag/AgCl; the printing ink is a composite of particles, binder and additives, and its composition changes electron transferSourced
- A membrane has three roles, selectivity, diffusion limiting and antifouling, and they trade off against each otherSourced
- Interference arises not only from electroactive species but also from haematocrit and the selectivity of the enzyme. Regulators specify test concentrations per substance and require them to be verifiedSourced
13. Glossary
- Working electrode (WE)
- The electrode at which the electrochemical reaction of interest occurs and the current is measured.
- Reference electrode (RE)
- The electrode that provides the potential reference. Usually Ag/AgCl, and it carries almost no current.
- Counter electrode (CE)
- The electrode that provides the return path for the current flowing at the working electrode.
- Potentiostat
- An instrument that measures current while controlling the working electrode's potential.
- Amperometry
- Measuring the current from oxidation or reduction at a constant potential.
- Voltammetry
- Measuring current while sweeping the potential.
- Potentiometry
- Measuring the potential difference between electrodes with almost no current, to find ion activity.
- Nernst equation
- The relation between potential and ion activity (concentration). About 59 mV per decade for a monovalent ion at 25 °C.
- Cottrell equation
- The equation showing that a diffusion-controlled current at a planar electrode decays in inverse proportion to √t.
- Mediator
- A small molecule that carries electrons between the enzyme and the electrode.
- GOx
- Glucose oxidase. Produces hydrogen peroxide from glucose and oxygen.
- Exclusion (permselective) membrane
- A membrane that stops interferents from arriving, by size or charge.
- Biofouling
- Performance loss when proteins or cells adhere to a surface.
- Screen-printed electrode
- A disposable electrode made by printing conductive ink onto a substrate through a screen.
- Pseudo-reference electrode
- An Ag/AgCl or similar electrode without an internal solution. Its potential is stable only when sample conditions stay constant.
- Haematocrit
- The volume fraction of red blood cells in blood. A source of error in strip-based glucose measurement.
14. References
- Thévenot DR, Toth K, Durst RA, Wilson GS (IUPAC) "Electrochemical biosensors: recommended definitions and classification" Biosens Bioelectron 2001;16:121–131 (abstract) — https://pubmed.ncbi.nlm.nih.gov/11261847/
- Grieshaber D, MacKenzie R, Vörös J, Reimhult E "Electrochemical Biosensors – Sensor Principles and Architectures" Sensors 2008;8:1400–1458 — https://pmc.ncbi.nlm.nih.gov/articles/PMC3663003/
- Hayat A, Marty JL "Disposable Screen Printed Electrochemical Sensors: Tools for Environmental Monitoring" Sensors 2014;14:10432–10453 — https://pmc.ncbi.nlm.nih.gov/articles/PMC4118360/
- Shinwari MW, Zhitomirsky D, Deen IA, Selvaganapathy PR, Deen MJ, Landheer D "Microfabricated Reference Electrodes and their Biosensing Applications" Sensors 2010;10:1679–1715 — https://pmc.ncbi.nlm.nih.gov/articles/PMC3264446/
- Weltin A, Kieninger J, Urban GA "Microfabricated, amperometric, enzyme-based biosensors for in vivo applications" Anal Bioanal Chem 2016;408:4503–4521 — https://pmc.ncbi.nlm.nih.gov/articles/PMC4909808/
- Yoo EH, Lee SY "Glucose Biosensors: An Overview of Use in Clinical Practice" Sensors 2010;10:4558–4576 — https://pmc.ncbi.nlm.nih.gov/articles/PMC3292132/
- Zhu Z, Garcia-Gancedo L, Flewitt AJ, Xie H, Moussy F, Milne WI "A Critical Review of Glucose Biosensors Based on Carbon Nanomaterials: Carbon Nanotubes and Graphene" Sensors 2012;12:5996–6022 — https://pmc.ncbi.nlm.nih.gov/articles/PMC3386727/
- Wisniewski N, Reichert M "Methods for reducing biosensor membrane biofouling" Colloids Surf B Biointerfaces 2000;18:197–219 (abstract) — https://pubmed.ncbi.nlm.nih.gov/10915944/
- U.S. Food and Drug Administration (FDA) "Self-Monitoring Blood Glucose Test Systems for Over-the-Counter Use — Guidance for Industry and FDA Staff", 29 September 2020 — https://www.fda.gov/regulatory-information/search-fda-guidance-documents/self-monitoring-blood-glucose-test-systems-over-counter-use
- ISO "ISO 15197:2013 In vitro diagnostic test systems — Requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus" (official page) — https://www.iso.org/standard/54976.html
- Nguyen TTH, Nguyen CM, Huynh MA, Vu HH, Nguyen TK, Nguyen NT "Field effect transistor based wearable biosensors for healthcare monitoring" J Nanobiotechnology 2023;21:411 — https://pmc.ncbi.nlm.nih.gov/articles/PMC10629051/
15. Claim-to-source audit
| Claim in the text | Basis | Label |
|---|---|---|
| That IUPAC defines an electrochemical biosensor as a self-contained, integrated device with a biological recognition element held in direct spatial contact with an electrochemical transduction element; that devices needing extra steps are "bioanalytical systems" and single-use devices that cannot monitor continuously are "single-use biosensors"; that classification is by biological specificity mechanism or signal transduction; and the performance criteria (sensitivity, operational and linear range, detection and quantification limits, selectivity, response times, sample throughput, reproducibility, stability, lifetime) | Abstract of Thévenot et al. (IUPAC) Reference 1 https://pubmed.ncbi.nlm.nih.gov/11261847/ | Sourced |
| Clark's oxygen electrode of 1955–56, the 1962 report of the first glucose sensor, and the 1975 commercial launch of a dedicated glucose biosensor; the classification of measurement methods (current, potential, conductance, impedance, FET); the roles of the three electrodes and the reference electrode usually being Ag/AgCl placed away from the reaction site; the use of platinum, gold, carbon and silicon compounds; the variables in the Cottrell and Nernst equations; ISE detection limits reaching 10⁻⁸ to 10⁻¹¹ M and the limited range of measurable ions; the definitions of enzyme-electrode generations, examples of mediators and the distance dependence of direct electron transfer; and the trend toward printed electrodes with low cost and ease of mass production | Grieshaber et al. Reference 2 https://pmc.ncbi.nlm.nih.gov/articles/PMC3663003/ | Sourced |
| That a screen-printed electrode is three electrodes on an inert substrate and the fabrication steps (substrate, working and auxiliary electrodes, reference, protective paste, modification); that inks consist of particles, polymer binder and additives with undisclosed formulations and that composition changes electron transfer; carbon's low background current, wide potential range and chemical inertness; the instability and reproducibility problems of mixing mediators into the paste; and that pre-anodising creates defect or edge-plane-like sites and oxygen functional groups, with overpotential reduction and peak separation reported in a comparison with oxygen-plasma treatment | Hayat and Marty Reference 3 https://pmc.ncbi.nlm.nih.gov/articles/PMC4118360/ | Sourced |
| That the Ag/AgCl reference electrode's standard potential is about 0.22 V at unit chloride activity; that AgCl dissolution shifts the potential and is worse for small solution volumes; that thin layers are used up and create mixed potentials; that diffusion barriers are used in microfabrication; the adhesion and paste-homogeneity issue for screen-printed paste; and that an Ag/AgCl electrode without internal solution is a pseudo-reference constant only in a particular sample | Shinwari et al. Reference 4 https://pmc.ncbi.nlm.nih.gov/articles/PMC3264446/ | Sourced |
| That hydrogen peroxide is oxidised on platinum and similar metals in a two-electron process; permselective membranes (size and charge exclusion, electropolymerised phenylenediamine, Nafion); enzyme immobilisation with BSA and glutaraldehyde and the addition of Nafion or polyurethane; the trade-off between linear range and sensitivity and the sensitivity cost of selective membranes; background subtraction with an enzyme-free electrode; interference removal with ascorbate oxidase; thin-film constructions on polyimide, ceramic and silicon with polyimide, SU-8 and nitride insulation; the course of protein adsorption (seconds to hours), foreign-body reaction (days) and encapsulation in vivo; passive coating and nitric oxide release; polyimide and similar materials having a modulus orders of magnitude above tissue; and the advice to measure interferent sensitivity on the bare electrode first | Weltin et al. Reference 5 https://pmc.ncbi.nlm.nih.gov/articles/PMC4909808/ | Sourced |
| That an electrochemical interferent is a species electroactive at the amperometric working potential; the haematocrit effects (red-cell oxygen competing with the mediator, viscosity slowing diffusion); that GDH-PQQ oxidises other sugars such as maltose and that overestimation occurs with icodextrin in peritoneal dialysis; that acetaminophen is oxidised directly after passing a membrane; and that FDA documents refer to ISO 15197 and CLSI EP-7A | Yoo and Lee Reference 6 https://pmc.ncbi.nlm.nih.gov/articles/PMC3292132/ | Sourced |
| That the first generation needs a relatively high potential to measure hydrogen peroxide so that ascorbic acid, uric acid, acetaminophen and others interfere; and that the countermeasures are selective membranes (Nafion, cellulose acetate) and a lower working potential (0–0.2 V vs Ag/AgCl) | Zhu et al. Reference 7 https://pmc.ncbi.nlm.nih.gov/articles/PMC3386727/ | Sourced |
| That nine kinds of modification have been examined to reduce biofouling: hydrogels, phospholipid-based biomimicry, flow-based systems, Nafion, surfactants, naturally derived materials, covalent attachment, diamond-like carbon and surface topology | Abstract of Wisniewski and Reichert Reference 8 https://pubmed.ncbi.nlm.nih.gov/10915944/ | Sourced |
| The interference evaluation in the FDA guidance of 29 September 2020 (for 510(k)s of OTC self-monitoring blood glucose systems; non-binding recommendations): three glucose levels, testing at the highest concentration with stepwise dilution, at least three lots with 30 replicates and relative bias with 95% confidence intervals, and labelling of known interfering conditions; and the Table 3 test concentrations (acetaminophen 20 mg/dL and others) and 26 items | FDA guidance Reference 9 https://www.fda.gov/regulatory-information/search-fda-guidance-documents/self-monitoring-blood-glucose-test-systems-over-counter-use | Sourced |
| That ISO 15197:2013 is the second edition (published May 2013), covers requirements, design-verification procedures and validation by intended users for in vitro blood-glucose monitoring systems measuring capillary blood glucose, is aimed at self-testing by lay persons at home, and was reviewed and confirmed in 2018 | ISO official page Reference 10 https://www.iso.org/standard/54976.html | Sourced |
| The in vitro slopes (−57, −48, −26 mV per decade) of an ISFET platform for Na⁺, K⁺ and Ca²⁺ in sweat, and the pH sensitivity of 58 mV/pH, as reported | Nguyen et al. (a report cited in the review) Reference 11 https://pmc.ncbi.nlm.nih.gov/articles/PMC10629051/ | Sourced |
| 1 kΩ × 10 µA = 10 mV; Cottrell currents (about 10.2 µA at 1 s, 3.2 µA at 10 s, 1.9 µA at 30 s); Nernst slopes (59.2 / 29.6 mV per decade at 25 °C, 61.5 mV per decade at 37 °C); 5.55 mM, 0.5 µL, about 2.8 nmol, about 540 µC; and a difference of about 4% between 25 °C and 37 °C | Our calculation. Assumes a 3 mm disc electrode, 1 mM, D = 7×10⁻⁶ cm²/s, n = 1, standard values of R, T and F, and 100 mg/dL = about 5.55 mM (molar mass 180.16) | Our calculation |
| The grouping of electrode materials into strengths and points of care; the possibility that the binder makes the active area differ from the apparent area; the three conflicting jobs of a membrane; the reading that passive coatings delay rather than eliminate fouling; the "what matters" notes in the layer-stack figure; thin film tending to cost more in equipment; and the adhesion issue of thin gold films | Commentary by this article. A materials-viewpoint rearrangement of the sources (partly marked as inference) | Commentary |
| The full text of the IUPAC recommendation, the numerical criteria of ISO 15197:2013, the current edition of the CLSI interference-testing document, specific ink formulations, real products' layer stacks, and long-term performance in or on the body | Not confirmable from published primary sources within the scope of this survey, so not stated | Commentary |
| That Figs. 1, 2, 3, 4 and 6 are explanatory drawings and not real measurement data or product structures, and that the hero image and Fig. 5 are AI-generated images | Note by this article | Commentary |
Last updated October 2026. Sources are limited to primary material (abstracts of IUPAC recommendation papers, peer-reviewed reviews, an FDA guidance document and an ISO official page). Because the article includes materials-design readings, those are marked as Commentary and kept separate from sourced fact. The full text of the IUPAC recommendation, the acceptance criteria of ISO 15197:2013, printing-ink formulations and the layer structure of real products are not stated here because no published primary source could be confirmed. The calculations in this article are derived from equations and illustrative assumptions and are not measurements from real sensors. All figures are explanatory concept graphics. Figs. 1, 2, 3, 4 and 6 are vector drawings; the hero image and Fig. 5 are AI-generated images, and none of them shows a real cross-section, micrograph or physical product. This article explains materials, principles, evaluation and regulation and is not medical advice.