MENU

Dual-line LFIA for AFP and CEA: ML-assisted screening specs

Research Square
Overview
A lightweight machine learning-assisted dual-line lateral flow immunoassay (LFIA) enables simultaneous on-site screening of AFP and CEA biomarkers. The system achieves a classification accuracy of 0.91 for both analytes using a KNN classifier, providing a semi-quantitative basis for clinical screening.
In Depth

Key Findings

The dual-line LFIA system demonstrates high analytical specificity and accuracy in the simultaneous detection of AFP and CEA. Using a KNN classifier with background-corrected RGB features, the system achieved an overall classification accuracy of 0.91 for both CEA and AFP. For CEA, the macro-averaged precision, recall, and F1-score were 0.92, 0.90, and 0.90, respectively; for AFP, these values were 0.93, 0.92, and 0.91. Quantitative analysis via Gaussian Process Regression (GPR) models yielded an R² of 0.932 for CEA (MSE = 84.455) and 0.902 for AFP (MSE = 121.506) on an independent test set.

Technical Details

The system utilizes antibody-conjugated gold nanoparticles (AuNPs) as probes. Transmission electron microscopy (TEM) confirmed that antibody conjugation did not substantially alter the oval morphology or colloidal stability of the AuNPs, though UV–vis absorption spectra showed a slight red shift in the localized surface plasmon resonance peak. Optimization determined the ideal antibody loading amount to be 6 µg mL⁻¹ for both AFP and CEA, with signal plateaus reached after approximately 10 min.

  • Feature Extraction: Background-corrected RGB features were selected over T/C intensity ratios and RGB feature intensity due to superior robustness against illumination variation.
  • Repeatability: For CEA, the coefficient of variation (CV) decreased from 26.3% at 10 ng mL⁻¹ to 0.13% at 100 ng mL⁻¹. For AFP, CV values were 11.15%, 7.06%, and 3.40% at 10, 50, and 100 ng mL⁻¹, respectively.
  • Recoveries: CEA recoveries were 115.0%, 84.7%, and 78.4% across the tested concentrations; AFP recoveries were 126.2%, 109.0%, and 72.4%.

Significance and Outlook

The integration of a smartphone-machine-learning workflow improves the interpretation of LFIA results at both interval and continuous concentration levels. By overcoming the limitations of visual reading, the GPR and KNN models provide a reliable semi-quantitative basis for screening-oriented decisions, reducing the impact of background heterogeneity and illumination variation in on-site environments.

Source: https://doi.org/10.21203/rs.3.rs-11475122/v1

Get our weekly technology intelligence — free

Receive an infographic that lets you judge at a glance whether each field’s analysis report is worth reading.

Subscribe Free — Weekly Tech Intelligence

By subscribing, you’ll receive Troy-Technical’s weekly technology intelligence newsletter.

  • Your email and selected fields are used only to deliver the newsletter.
  • We never share your information with third parties.
  • You can unsubscribe anytime via the link in each email.

See our Privacy Policy for details.

Takes about a minute · Unsubscribe anytime

Published by Troy-Technical, an independent site run by one engineer with a career in materials development.
About the author / Contact info@troy-technical.jp
Let's share this post !

Author of this article

TOC